Brookfield CT3 Texture Analyzer Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
Brookeld Engineering Labs., Inc. Page 24 Manual No. M08-372-C0113
Duringthetesttherunningscreenisseenasshownbelow:
TEST: TENSION
DEFORMATION:XXXX.X mm
LOAD: XXXX.X g
III.25
Deformation and Load will remain live displays during the test. When deformation reaches the
target value, the test stops.
Testresultsareshownonthereportingscreen:
PEAK LOAD: XXX.X g
DEF@PEAK: XXX.X mm
WORK: XXXX.X mJ
FINAL LOAD: XXXX.X g
TEMP: XXX.X F
III.26
Peak load is the maximum load measured during the test. Temperature at peak load is only shown
if temperature probe is used.
Def@Peak is the sample deformation at the peak load. The trigger position is the zero deformation
reference point.
Workistheareaundertheloadcurveandismeasuredinmillijoules.
Final load is the load at the target deformation.
NOTE:Itisnotpossibletoperformtensiontestswiththe100gloadcell.
SURIMI TEST - performs a single compression of a surimi sample until the gel ruptures. The
testresultwillbethedistancethesampledeformedbeforeitrupturedandthepeakloadjustprior
to the rupture. As soon as the rupture occurs the probe retracts to the HOME starting position.
NOTE: ThistestisonlyavailableinCT3instrumentswithloadcellsupto4500gandrm-
ware version 2.0 or higher.
The operator may set the trigger value, correction value and test speed.
TEST: SURIMI
TRIGGER: 5.0 g
CORRECTION: 30 g
SPEED: 1.0 cm/s
III.27
The trigger value is a small load, in grams, measured by the CT3 that indicates the probe is in contact