I.3 Specications
Range: 40–141 KU
32–1,099 gm
27 - 5,274 cP*
Accuracy: ± 1% of full scale range
Repeatability: ± 0.5% of full scale range
Paddle speed: 200 rpm ± 0.1 rpm
Printer output: 25D-Type Centronics Parallel Printer Output
Net weight: 22 lb. (10 kg)
Gross weight: 25 lb. (11 kg)
Dimensions: 15 x 11 x 20
Operating Environment: 0°C (32°F) to 40°C (104°F)
20% - 80% R.H.: non-condensing atmosphere
Electrical Certications: Conforms to CE Standards for: Electromagnetic Compatibility
(EMC), Low Voltage (LVD) and Safety Requirements for electrical equipment for measurement
control and laboratory use.
*Centipoise scale is for reference only. Do not use for calibration.
I.4 Sample Container Specications
Container Dimensions
Can Height Bottom Flange Diameter
U.S. Pint 3.850" (9.78 cm) 3.385" (8.60 cm)
U.S. Half Pint 2.850" (7.24 cm) 2.875" (7.30 cm)
U.S. Quart 4.845" (12.31 cm) 4.230" (10.74 cm)
If you have a requirement for container dimensions other than those listed above, please contact
Brookeld or an authorized Brookeld agent.
I.5 Utilities
Input voltage: 115 VAC or 230 VAC*
Input frequency: 50/60 Hz
Power consumption: 16 VA (watts)
Power cord color code:
USA Outside USA
Hot (live) Black Brown
Neutral White Blue
Ground (earth) Green Green/Yellow
*Main supply voltage uctuations are not to exceed ± 10% of the nominal supply voltage.