Brookeld Engineering Labs., Inc. Page 51 Manual No. M02-215-B0412
where:
θ
cal
= Spring windup angle (revolutions)
V
cal
= Calibrationspeed(xedat0.1rpm) (rpm)
t
cal
= Calibration time (milliseconds)
(time for 0% to 100% spring wind up)
t
cal
=
bi
x
V
x
100
b
cal
V
cal
where:
t
cal
= Calibration time (milliseconds)
(time for 0% to 100% spring windup
during calibration)
bi = Base increment (milliseconds)
b
cal
= base increment calibration torque (%torque/base increment)
V = Speed (rpm)
V
cal
= Calibration speed (rpm)
(xedat0.1rpm)