Brookeld Engineering Labs., Inc. Page 5 Manual No. M02-312A0607
I.3 Specications
Torque Range: Low 797-7,970 dyne•cm (designated on serial tag as 1/23 CAP)
High 18,100-181,000 dyne•cm (designated on serial tag as 1.0 CAP)
Speeds: 750 RPM and 900 RPM or
single speed from 5-900 RPM as specied at time of order
Temperatures: CAP 1000+L 5°C (or 15°C below ambient, whichever is higher) to
75°C
CAP 1000+H 50°C to 235°C
All models provide 0.1°C increments
Weight: Gross Weight 36 lb 16.3 kg
Net Weight 27 lb 12.3 kg
Carton Volume 4.9 cu ft 0.15 m
3
Carton Dimensions 18 in. L x 18 in. W x 26 in. H
48 cm. L x 48 cm. W x 66 cm. H
Materials: CAP cone spindles and temperature plates are made of tungsten carbide.
Operating CAP 1000+ Viscometers must be operated within the following
Environment: ambient temperatures: +5°C (41°F) to 40°C (104°F)
and humidity: 20% to 80% R.H. (non-condensing atmosphere)
I.4 Installation
Note: DO NOT lift the viscometer by the handle or head! LIFT only by the base console or
column!
1) Set the viscometer on a clean level bench surface.
2) Remove shipping foam insert from the plate area on the CAP Viscometer. Store the
foam insert for future use when shipping or transporting CAP Viscometer.
Foam Insert Used
When Shipping
CAP Viscometer
Figure I-2: Detail of Foam Insert