Brookeld Engineering Labs., Inc. Page 90 Manual No. M13-167-B0614
Viscometer Will Not Return to Zero
Viscometerisnotlevel
• Checkwithspindleoutofthesample.
• Adjustthelaboratorystand.
Pivotpointorjewelbearingfaulty
• PerformanOscillationCheck*
RemovethespindleandturnthemotorOFF;selectdisplayto%torquemode.
Gentlypushupontheviscometercoupling.
Turnthecouplinguntilthedigitaldisplayreads10-15onthe%display.
Gentlyletgoofthecoupling.
Watchthedigitaldisplay;youshouldseea“run”ofnumbersnexttothe%;the
“run”ofnumbers shouldultimatelystopat0.0(+/-0.1).
IfthedigitaldisplaydoesnotreturntoZERO,theunitmostlikelyisinneedof
service.
• Performcalibrationcheck(seeAppendixF).
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerfor
repair(seeAppendixO).
*Thisproceduredoesnotapplytoinstrumentswithballbearingsuspension(seeSection
I.3).
Display Reading Will Not Stabilize
Specialcharacteristicofsampleuid.Thereisnoproblemwiththeviscometer.
• RefertoAppendixC.
Checkforerraticspindlerotation.
• Verifypowersupply
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerfor
repair.
Bent spindle or spindle coupling.
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerfor
repair.
Temperatureuctuationinsampleuid.
• Usetemperaturebathforcontrol.