Brookeld Engineering Labs., Inc. Page 12 Manual No. M/05-370-B0507
Once the test mode is selected the test parameters must be set.
•Normalrequires: Trigger,Distance,Speed.
•HoldTimerequires: Time,Trigger,Distance,Speed.
•CycleCountrequires: Count,Trigger,Distance,Speed.
•Bloomisaxedtestwithpresetparameters.
•Static Load is for verifying calibration and does not require any parameters.
Thetestparametersaredenedasfollows:
Trigger: The load, in grams, measured by the LFRA to indicate that the probe is in con-
tact with the sample. Once the trigger value is reached, the test will begin at the
denedspeed.Brookeldrecommendsatriggervalueof0.5%oftheloadcell
range. For example, use 5g trigger with a 1000g load cell instrument.
Distance: The total downward distance the probe will travel once the trigger value is
reached.
Speed: Thespeedatwhichtheprobewilltravelthespecieddistance.
Time: Thenumberofsecondsthattheprobewillbeheldatthedeneddistanceduring
a Hold Time test.
Count: The number of cycles (Speed and Distance) that will be applied to the sample
during a Cycle Count test.
The parameters must be set within the ranges shown in Table III.1.
Parameter 100g 1000g 1500g 4500g
Trigger 0.1 – 10g 0.2 – 100g 0.2 – 150g 0.6 – 500g
Distance 0.1 – 75mm 0.1– 75mm 0.1 – 75mm 0.1 – 75mm
Speed 0.1 – 10 mm/s 0.1 – 10 mm/s 0.1 – 10 mm/s 0.1 – 10 mm/s
Hold Time 0 – 9999 s 0 – 9999 s 0 – 9999 s 0 – 9999 s
Cycle Count 0 – 99 0 – 99 0 – 99 0 – 99
Table III.1
Deneatestbyrstselectingatestmode.RotatetheSelect/Scrollknobuntiltherequiredtest
modeisdisplayed,thenpresstheSelect/Scrollknobtoconrmyourchoice.Asthetestmodes
aredisplayed,theparametersspecictothatmodewillbeshownwiththepreviouslyselected
parameters.
Onceatestmodeisselected,thecursorwillmovetotherstparameter.Youmayeitherselect
that parameter to enter new values by pressing the Select / Scroll knob or you may scroll to the
next parameter by rotating the Select / Scroll knob.