Brookeld Engineering Laboratories, Inc. Page 23 Manual No. M/07-022-D0613
%=76.4 RPM=100
TIMED STOP DONE
Figure II-22
The display will switch between that of Figures II-21 and II-22 for each press of either the
UP or DOWN ARROW key.
The test can be repeated by pressing the motor ON/OFF key.
Pressing any key except the UP or DOWN ARROW keys or motor ON/OFF will cause the
viscometer to exit the Timed Stop mode and resume normal operation.
NOTE: For the Timed Stop method, the DV-I PRIME Viscometer will retain the last
value for the time interval in MEMORY so that it will become the default
the next time the user elects to use this method.
II.9.3 Time to Torque
1. After pressing the
UP OR DOWN key when in the display of Figure II-16, the user selects
time to torque when ashing and presses enter.
The following screen will be presented:
TIMED TORQUE
SET TORQUE:00%
Figure II-23
Using the
UP and DOWN ARROW keys, the user enters a value for the torque level, which
the viscometer must achieve. This value can be as high as 99%.
2. The user presses the
ENTER key one more time to end the torque input at which point the
viscometer will display the following screen:
TIMED TORQ:PRESS
MOTOR ON/OFF
Figure II-24
3. At this point, the user need only press the
MOTOR ON/OFF key to begin the timed torque
operation.
4. We will assume that the user pressed the
MOTOR ON/OFF key to ON and is now presented
with the following display for the duration of the timed torque run: