Brookeld Engineering Laboratories, Inc. Page 54 Manual No. M/07-022-D0613
Table D-2 lists the model codes and spring torque constants for each viscometer model.
VISCOMETER MODEL
TORQUE
CONSTANT TK
MODEL CODE ON DV-I
PRIME SCREEN
LVDV-I PRIME 0.09373 LV
2.5xLVDV-I PRIME 0.2343 2.5LV
5xLVDV-I PRIME 0.4686 5LV
1/4 RVDV-I PRIME 0.25 1/4RV
1/2 RVDV-I PRIME 0.5 1/2RV
RVDV-I PRIME 1 RV
HADV-I PRIME 2 HA
2xHADV-I PRIME 4 2HA
2.5xHADV-I PRIME 5 2.5HA
HBDV-I PRIME 8 HB
2xHBDV-I PRIME 16 2HB
2.5xHBDV-I PRIME 20 2.5HB
Table D-2
The full scale viscosity range for any DV-I Prime model and spindle may be calculated using the
equation:
Full Scale Viscosity Range [cP] = TK
*
SMC
*
10,000
RPM
where:
TK = DV-I Prime Torque Constant from Table D-2
SMC = Spindle Multiplier Constant from Table D-1
The Shear Rate calculation is:
Shear Rate (
1
/
Sec
) = SRC
*
RPM
where:
SRC = Shear Rate Constant from Table D-1
Comentarios a estos manuales