Brookeld Engineering Laboratories, Inc. Page 46 Manual No. M/07-022-D0613
A.5 Verifying Calibration
1. Determine the appropriate sample volume. Refer to Table A-1 to determine the correct sample
volume required for the spindle to be utilized.
2. Select a Brookeld Viscosity Standard uid that will give viscosity readings between 10% and
100% of Full Scale Range. Refer to Appendix B for viscosity ranges of cone spindles; ranges
listed apply to CPE cones.
Do not use a silicone viscosity standard uid with a viscosity value greater than 5000 cP with a
Cone/Plate. Brookeld offers a complete range of mineral oil viscosity standards suitable for
use with Cone/Plates for viscosities above 5,000 cP or shear rates above 500 sec
-1
; see Table E-1
and E-2 in Appendix E for a list of available uids.
It is best to use a viscosity standard uid that will be close to the maximum viscosity for a given
cone spindle/speed combination.
Example: LVDV-I PRIME Viscometer, Cone Spindle CPE-42, Brookeld Silicone
Viscosity Standard having a viscosity of 9.7 cP at 25°C.
At 60 RPM, the Full Scale Viscosity Range is 10.0 cP. Thus, the viscometer reading should be
97% torque and 9.7 cP viscosity ± 0.197 (closer to ±0.2) cP. The accuracy is a combination of
viscometer and uid tolerance (refer to Interpretation of Calibration Test Results at the end
of Appendix E).
3. With the motor off, remove the sample cup and place the viscosity standard uid into the cup.
Cone Part No.
Sample Volume
CPE-40
0.5 mL
CPE-41
2.0 mL
CPE-42
1.0 mL
CPE-51
0.5 mL
CPE-52
0.5 mL
Table A-1
4. Attach the sample cup to the viscometer and allow sufcient time for the sample, cup and cone
to reach temperature equilibrium.
5. Turn the motor on. Set the desired speed(s). Measure the viscosity and record the reading in
both % torque and centipoise (cP).
NOTE: The cone spindle must rotate at least ve (5) times before a viscosity reading is
taken.
6. Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier, for
the specic viscosity standard uid(s) that you are using.
* The CPE designation on the cone spindle indicates use with Electronic Gap Setting Cone/Plate
Viscometers/Rheometers only.