Brookeld Engineering Labs., Inc. Page 20 Manual No. M13-167-B0614
Range
Figure II-6
TIP: The Rangevalue isthe same astheAutoRange availableon earlier Brookeld
viscometer models.
II.7 Out of Range
TheDV2TViscometerwillgiveonscreenindicationswhenthemeasurementisoutofrange
oftheinstrument.Whenthe%Torquereadingexceeds100%(overrange),thedisplay
of%Torque,Viscosity,andShearStresswillbeEEEE(seeFigureII-7).Ifthe%Torque
valueisbetween0-9.9%,thedataeldlabelwillash.Whenthe%Torqueisbelowzero
(negativevalues),thedisplayofViscosityandShearStresswillbe----.
TIP:Brookfield recommends collecting data only when the %Torque reading is
between10-100%.
Figure II-7
It's good information