BrookeldEngineeringLabs.,Inc. Page34 ManualNo.M08-218-B0212
The rst user dened value ( = the rst measuring point) is always the start value of the
ramp. The last measuring point is determined as the end value of the user dened value
range. As in our example, to reach the values D = 10, 20, 30, ... , 100 [s
-1
] a starting value
of 10 [
s-1
] is used with increments of 10, therefore, 10 measuring points is required. To
check: 100-10/(10-1) = 10.
Number of Measurement Points = ((End Value - Start Value)/Pre-set Value - 1)
After selection of the menu level “Edit Program”, the user will be prompted to select the
program.
Select program
1) Prog xyz
2) Test
3) NEW
Available programs
contain a name
All free (non-dened)
programs are marked NEW
All free programs are initially marked as NEW. To avoid overwriting existing programs,
select NEW as a program to be edited and select a measuring system. After the selection of
the program with the
OK
key, you will be requested to select a measuring system.
Attention: Any dened program MUST ALWAYS be executed with the same measuring
system, otherwise improper results will be calculated.
Select Meas. system:
1) C25-1/30 >
2) C25-2/30
3) C50-1/30 >
After selecting the measurement system, enter the number of steps (number of ramp and
straight line functions). The number of steps can range from 1 to a maximum of 10 steps.
In this example, we need three steps, so the number “01” is changed to “03” - (see Section
VI.4: Input of Numerical Values and Alphnumeric Texts). The message “Range error” will
appear if <1 or >10 is entered.
Nr of steps: 03
Input number of
steps 1 - 10. This
example is 03.
After entering the number of steps, input the type of measurement.
Select Input:
-D[1/s]
-n[rpm]
-Tau[Pa] >
Vista de pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 61 62

Comentarios a estos manuales

Sin comentarios

Allied Telesis Switch AT-9000/12POE manuals

Owner’s manuals and user’s guides for Network switches Allied Telesis Switch AT-9000/12POE.
We providing 1 pdf manuals Allied Telesis Switch AT-9000/12POE for download free by document types: User Manual






More products and manuals for Network switches Allied Telesis

Models Document Type
AT-8000GS/24 User Manual   Allied Telesis AT-8000GS/24 User's Manual, 52 pages
990-004270-50 Datasheet   Allied Telesis AT-GS924M, 2 pages
990-004132-50 Datasheet   Allied Telesis AT-x230-18GP, 6 pages
990-004277-50 Datasheet   Allied Telesis AT-FS917M, 2 pages
Patch SB244-01 User Manual   Allied Telesis Patch SB244-01 User's Manual, 4 pages
AT-9000/24 User Manual   Allied Telesis at-9000/24 User's Manual, 326 pages
AR441S User Manual    Allied Telesis AR441S User's Manual, 54 pages
AT-AR750S-DP User Manual   Allied Telesis AT-AR750S-dp User's Manual, 5 pages
990-004272-50 Datasheet   Allied Telesis AT-x510L-28GP, 9 pages
AT-9408LC/SP User Manual   Allied Telesis AT-9408LC/SP User's Manual, 84 pages
Switch AT-S95 User Manual   Allied Telesis Switch AT-S95 User's Manual, 11 pages
990-003857-50 Datasheet   Allied Telesis AT-x210-24GT, 6 pages
990-004031-50 Datasheet   Allied Telesis AT-x230-10GP, 6 pages
990-003944-50 Datasheet   Allied Telesis AT-x310-50FT, 6 pages
990-002653-50 Datasheet   Allied Telesis AT-9000/12POE, 6 pages
AT-X210-9GT Datasheet   Allied Telesis AT-X210-24GT network switch, 6 pages
990-003848-00 Datasheet   Allied Telesis AT-x510-28GTX, 8 pages
AT-GS900/8PS Datasheet   Allied Telesis AT-GS900/8PS, 2 pages
AT-IX5-28GPX Datasheet   Allied Telesis AT-IX5-28GPX network switch, 7 pages
AT-X510-28GTX-50 Datasheet   Allied Telesis AT-x510-52GPX, 8 pages