BrookeldEngineeringLabs.,Inc. Page52 ManualNo.M08-218-B0212
The AC adapter unit of the R/S-CPS+ Rheometer permits operation of the rheometer with supply
voltages ranging from 100 to 240 VAC with frequencies from 50 to 60 Hz.
The main connecting cable supplied with the R/S-CPS+ Rheometer may not, in some cases, meet
the requirements of every country. It is absolutely necessary that you use a main connecting cable
that meets the specic requirements and regulations of the relevant country.
The following information explains the requirements that are to be taken into account when
choosing the main connecting cable.
Generalinformation
The connecting cable must be permitted (authorized) in the country where it is used.
The mains cable should be at least 2 m and at the most 3 m long.
USAandCanada
UL-permission and CSA Certicate are required for the mains connecting cable.
The following minimal requirements are valid for the cable:
- No. 18 AWG
- Type SV or similar
- 3-phase
The cable must have a nominal rating of at least 10 A.
The main plug of type NEMA 5-15P (15A, 125V) or NEMA 6-15P (15A, 250V) must
have grounding contact.
Japan
The following cable types and connection values are required in Japan:
All parts of the cables (cable, socket and plug) must have brand name and registration
number according to the Dentori law.
The following minimal requirements apply to the cable:
- 0.75 mm2, 2-phase
- Type VCT or VCTF
- 3-phase
The cable must have a minimum nominal rating of 7 A.
The mains plug must be a 2-pin plug with grounding contact according to the Japanese
Industrial Standard C8303 (15A, 125V).