Brookeld Engineering Labs., Inc. Page 30 Manual No. M13-2100
III.5 Temperature Control
TheDV3TRheometerprovidestheabilitytocontrolthetemperatureofaconnectedBrookeld
temperaturecontroldevicesuchastheThermosel(HT-106controller)andwaterbathswithSDor
APcontrollers(forexample:TC-550SD).TheThermoselorwaterbathcanthenbeusedtocontrol
thetemperatureofthesampleundertest.
TheDV3Tconnectstothetemperaturecontroldevicethroughadedicatedcommunicationcable.
TheThermoselrequirescableDVP-141.ThewaterbathsrequirecableDVP-207.Thesecables
areavailablefromBrookeld.
Whenconguredforcontrolofatemperaturecontroldevice,theDV3Tmayhavetwotemperature
inputs;
1)thetemperatureprobesuppliedwiththeDV3T(DVP-94Y,CPA-44PYZ,orSC4-13RPY)
and
2)thetemperatureprobesuppliedwiththetemperaturecontroldevice(waterbathorThermosel).
Onlyonetemperatureprobecanbeusedfordisplayoflivetemperature,control,anddatacollection.
SelecttheprobetobeusedintheTemperaturemenu(seeFigureIII-6).
TIP:EmbeddedtemperatureprobesusedwiththeCone/PlateandSmallSampleAdaptercanbe
usedforsetpointcontrolofawaterbath.Makingthisselectioncanimprovethetemperature
controlofthesamplebyoffsettingthetemperaturebathtoaccountforanyheatlossin
thetubingorwaterjacket.Tousetheembeddedtemperatureprobe,selectProbeinthe
Temperaturemenu.
TemperaturecontrolmustbeinitiatedeachtimetheDV3Tispoweredon.Besurethatthetemperature
controldeviceispoweredonpriortoinitiatingcommunication.ChoseDeviceSetupintheSettings
menu.ChoseTemperatureandyouwillbepresentedwiththeTemperaturemenu(seeFigureIII-6).
TemperatureOffsetisdescribedinSectionIV.1.
Figure III-6
SelectTemperatureControllerandyouwillbepresentedwiththeDefaultTemperaturemenu(see
FigureIII-7).Thismenuallowstheuserto1)initiatecommunicationwiththetemperaturecontroller
and2)setadefaulttemperature.