Brookfield DV2+Pro Viscometer Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 101
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
Brookeld Engineering Labs., Inc. Page 42 Manual No. M03-165-F0612
where the operator is instructed to mount a spindle 31 and then press a key to continue. Or you
mightsee:
SET TEMP TO 100C
PRESS A KEY
Figure IV-42
followedby:
WAIT TEMP 100C
P1 S00/12 00:14
Figure IV-43
Here, the B.E.V.I.S. program is waiting for the temperature to reach one hundred (100) °C before
it continues to the next program step. Also displayed is the program number (P1), the step
number and the total program steps (S00/12) and the elapsed time since the program (or step)
began (00:14). While in this “wait state” you can press the OPTIONS/TAB key to see the viscometer
defaultscreeninFigureIV-44;thePGMwillbeashing.
CP 123.4 55.1C
10 RPM PGM% 63.7
Figure IV-44
The DVLOADER Software is used to create, save, print, and down load programs to the DV-
II+Pro Viscometer. The next section explains how to use the DVLOADER software.
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 100 101

Comentarios a estos manuales

Sin comentarios