Brookeld Engineering Labs., Inc. Page 96 Manual No. M03-165-F0612
Recorder Pen Moves in Wrong Direction
Output polarity reversed
• Reverseleads
Viscometer Wil Not Communicate with PC
Check the comm port and make sure the correct port is being utilized.
Check the interconnecting cable for proper installation
Check the Options menu and make sure the PC PROG is set to either “ON” or “OFF” in
accordance with the operating instructions for the program/procedure in use.
Iftheabovedonotrectifytheproblem,dothefollowing:
Shut off viscometer
Attach printer.
Press and hold the MOTOR ON/OFF and ENTER/AUTORANGE keys simulanteously while turning
the viscometer power on. Figure K-1 appears on the viscometer display.
VISCOMETER SETUP
ENTER TO START
Figure K-1
Press ENTER/AUTORANGEkeyandFigureJ-2appears.
PRINT PARAM RAM?
NO#$THEN ENTER
PRINT PARAM RAM?
YES#$THEN ENTER
"
t
Figure K-2
ScrolltoYESandpresstheENTER/AUTORANGE key. Figure K-3 appears.
READY PRINTER
ENTER TO START
Figure K-3
Press ENTER/AUTORANGE. Information similar to Figure K-4 will appear on your printer
output.