Brookeld Engineering Labs., Inc. Page 96 Manual No. M03-165-F0612
❏ Recorder Pen Moves in Wrong Direction
✓ Output polarity reversed
• Reverseleads
❏ Viscometer Wil Not Communicate with PC
✓ Check the comm port and make sure the correct port is being utilized.
✓ Check the interconnecting cable for proper installation
✓ Check the Options menu and make sure the PC PROG is set to either “ON” or “OFF” in
accordance with the operating instructions for the program/procedure in use.
Iftheabovedonotrectifytheproblem,dothefollowing:
✓ Shut off viscometer
✓ Attach printer.
✓ Press and hold the MOTOR ON/OFF and ENTER/AUTORANGE keys simulanteously while turning
the viscometer power on. Figure K-1 appears on the viscometer display.
VISCOMETER SETUP
ENTER TO START
Figure K-1
✓ Press ENTER/AUTORANGEkeyandFigureJ-2appears.
PRINT PARAM RAM?
NO#$THEN ENTER
PRINT PARAM RAM?
YES#$THEN ENTER
"
t
Figure K-2
✓ ScrolltoYESandpresstheENTER/AUTORANGE key. Figure K-3 appears.
READY PRINTER
ENTER TO START
Figure K-3
✓ Press ENTER/AUTORANGE. Information similar to Figure K-4 will appear on your printer
output.