Brookeld Engineering Labs., Inc. Page 67 Manual No. M03-165-F0612
A.3 Setting the Gap
1. Move the toggle switch to the right; this will
turn on (enable) the Gap Setting Feature. The
Pilot (red) light will be illuminated.
Note: The motor should be OFF.
2. If the contact light (yellow) is illuminated, turn
the micrometer adjustment ring clockwise (as
you look down on the instrument) until the light
is no longer illuminated (see Figure A-5).
3. If the yellow contact light is not illuminated,
slowly turn the micrometer adjustment ring in
small increments (one or two scale divisions)
counter-clockwise.
Continue moving the micrometer adjustment ring
slowly counter-clockwise until the contact light
(yellow)rstturns on.THIS IS THE “HIT
POINT.”
4. Adjust the sliding reference marker, right or
left, to the closest full scale division mark (see
Figure A-6).
5. Turn the micrometer adjustment ring one scale
division to the left to meet the line on the sliding
reference marker. THE YELLOW CONTACT
LIGHT SHOULD GO OFF.
6. Youhaveestablishedthegapspaceneededfor
measurement. Now turn the toggle switch OFF
(left); the red pilot light should go off.
Theviscosityofelectricallyconductiveuids
may be affected if readings are taken with the
Electronic Gap Setting feature “on”. Be sure to
shut the feature “off” before taking readings!
7. Carefully remove the sample cup.
Note:
1. The cup may be removed and replaced without resetting the gap if the micrometer
adjustment ring has not been moved.
2. Remove the spindle from the viscometer when cleaning.
3. Re-establish the hit point every time the spindle is attached/detached.
Moves Away
from Hit Point
(clockwise)
LEFTx
Moves Towards
Hit Point
(counter-clockwise)
RIGHT
Full Scale
Division Marks
Sliding
Reference
Marker
Figure A-5
Figure A-6
Vista de pagina 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 100 101

Comentarios a estos manuales

Sin comentarios