Brookeld Engineering Labs., Inc. Page 68 Manual No. M03-165-F0612
A.4 Verifying Calibration
1. Determine the appropriate sample volume. Refer to Table A-1 to determine the correct sample
volume required for the spindle to be utilized.
2. SelectaBrookeldViscosityStandarduidthatwillgiveviscosityreadingsbetween10%and
100% of full scale range. Refer to Appendix B for viscosity ranges of cone spindles.
BrookeldusesmineraloilviscositystandarduidstocalibrateWellsBrookeldCone/Plate
Viscometersatthefactory.Brookeldrecommendsthatcustomersusemineraloilviscosity
standarduidswhenyouperformacalibrationcheck.
Ifyoudecidetouseasiliconeviscositystandarduid,donotuseauidwithaviscosityvalue
greaterthan5000cPwithaCone/Plate.Brookeldoffersacompleterangeofmineraloilviscosity
standards suitable for use with Cone/Plates for viscosities above 5,000 cP or shear rates above
500 sec
-1
;seeTableE-2inAppendixEforalistofavailableuids.
Itisbesttouseaviscositystandarduidthatwillbeclosetothemaximumviscosityforagiven
cone spindle/speed combination.
Example: LVDV-II+ Pro Viscometer, Cone Spindle CPA-42Z, Brookeld Silicone Viscosity
Standard having a viscosity of 9.7 cP at 25°C
At 60 RPM, the full scale viscosity range is 10.0 cP. Thus, the Viscometer reading should be
97%torqueand9.7cPviscosity±0.197cP.Theallowableerror(±0.197cP)isacombination
ofViscometeraccuracyanduidtolerance(refertoInterpretation of Calibration Test Results
in Appendix E).
3. Withthemotoroff,removethesamplecupandplacetheviscositystandarduidintothecup.
Table A-1
Cone Part No.
CPA-40Z, CPE-40, CP-40
CPA-41Z, CPE-41, CP-41
CPA-42Z, CPE-42, CP-42
CPA-51Z, CPE-51, CP-51
CPA-52Z, CPE-52, CP-52
Sample Volume
0.5 mL
2.0 mL
1.0 mL
0.5 mL
0.5 mL
4. AttachthesamplecuptotheViscometerandallowsufcienttimeforthesample,cupandcone
to reach temperature equilibrium.
5. Turn the motor on. Set the desired speed(s). Measure the viscosity and record the reading in
both % torque and centipoise (cP).
6. Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier, for
thespecicviscositystandarduid(s)thatyouareusing.
The CPE or CPA designation on the cone spindle indicates use with Electronic Gap Setting
Cone/Plate Viscometers/Rheometers only.