Brookeld Engineering Laboratories, Inc. Page 114 Manual No. M/98-211-B0104
Brookeld Engineering Laboratories, Inc. Page 114 Manual No. M98-211-E0912
APPENDIX E - Calibration Check Procedures
Brookeld’s accuracy statement for viscometers used with standard spindles is +/- 1% of full scale
range. When measuring viscosity with a specic spindle rotating at a dened speed, the maximum
viscosity that can be measured is dened as full scale range. For digital viscometers, this value is
easily determined by pressing the “AUTORANGE” key. The display will how the full scale range
viscosity in cP or mPa•s and the torque value will show “100%”. Multiply the full scale range vis-
cosity by 1% to determine the accuracy of any future measurement made with that spindle/speed
combination.
When using the following accessory devices with your viscometer, the accuracy is +/- 2%. Dimen-
sional tolerances in the accessory device allow for the increase from +/- 1% to +/-2%.
Small Sample Adapter
Thermosel
Ul Adapter
DIN Adapter
Spiral Adapter
For more help you can go to the website,
www.brookeldengineering.com, and download the video.
The accuracy of the DV-III Ultra is veried using viscosity standard uids which are available from
Brookeld Engineering Laboratories or your local Brookeld agent. Viscosity standards are Newtonian,
and therefore, have the same viscosity regardless of spindle speed (or shear rate). Viscosity standards,
calibrated at 25
°C, are shown in Table E-1.
Container size: For Viscosity Standards < 30,000 cP, use a 600 ml Low Form Grifn Beaker
having a working volume of 500 ml.
For Viscosity Standards
30,000 cP, use the uid container.
Inside Diameter: 3.25" (8.25cm)
Height: 4.75" (12.1cm)
Note: Container may be larger, but may not be smaller.
Temperature: As stated on the uid standard label: (
+/-) 0.1°C
Conditions: The DV-III Ultra should be set according to the operating instructions. The water