Brookeld Engineering Laboratories, Inc. Page 129 Manual No. M/98-211-B0104
Brookeld Engineering Laboratories, Inc. Page 129 Manual No. M98-211-E0912
Viscometer Will Not Return To Zero
Rheometer is not level
Adjust the laboratory stand
Pivot point or jewel bearing faulty
Perform an Oscillation Check*
Manually lift the spindle nut and rotate counter clockwise until the display shows
20% - 30% deection.
Release the spindle nut.
The % display should return to 0.0% ± 0.1%.
Perform calibration check
Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
* This procedure does not apply to instruments with ball bearing suspension (See Section I.7).
Display Reading Will Not Stabilize
Special characteristic of sample uid. There is no problem with the rheometer.
Refer to Appendix C
Check for erratic spindle rotation
Verify power supply
Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
Bent spindle or spindle coupling.
Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
Temperature uctuation in sample uid
Use temperature bath control.
No Recorder Response
Be sure the rheometer is not at ZERO reading.
Be sure the recorder is ON and not on STANDBY.
Verify the range settings.
Check cable leads for clean connection.
Recorder Pen Moves In Wrong Direction
Output polarity reversed
Reverse leads