❑ Viscometer Will Not Return To Zero
✓ Rheometer is not level
• Adjust the laboratory stand
✓ Pivot point or jewel bearing faulty
• Perform an Oscillation Check*
✓ Manually lift the spindle nut and rotate counter clockwise until the display shows
20% - 30% deection.
✓ Release the spindle nut.
✓ The % display should return to 0.0% ± 0.1%.
• Perform calibration check
• Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
* This procedure does not apply to instruments with ball bearing suspension (See Section I.7).
❑ Display Reading Will Not Stabilize
✓ Special characteristic of sample uid. There is no problem with the rheometer.
• Refer to Appendix C
✓ Check for erratic spindle rotation
• Verify power supply
• Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
✓ Bent spindle or spindle coupling.
• Contact Brookeld Engineering Laboratories, Inc. or your Brookeld dealer for repair.
✓ Temperature uctuation in sample uid
• Use temperature bath control.
❑ No Recorder Response
✓ Be sure the rheometer is not at ZERO reading.
✓ Be sure the recorder is ON and not on STANDBY.
✓ Verify the range settings.
✓ Check cable leads for clean connection.
❑ Recorder Pen Moves In Wrong Direction
✓ Output polarity reversed
• Reverse leads