Brookeld Engineering Laboratories, Inc. Page 118 Manual No. M/98-211-B0104
Brookeld Engineering Laboratories, Inc. Page 118 Manual No. M98-211-E0912
Calibration Procedure using a Helipath Stand and T-Bar Spindles
When a Helipath Stand and T-Bar spindles are used:
Remove the T-bar spindle and select a standard LV(#1-3) or RV,HA,HB (#2-6) spindle. Follow the
procedures for LV(#1-3) and RV,HA,HB (#2-6) Brookeld spindles outlined above.
T-Bar spindles should not be used for verifying calibration of the DV-III Ultra Rheometer.
Calibration Procedure for Spiral Adapter
1) Place the viscosity standard uid (in the proper container) into the water bath.
2) Attach the spindle to the viscometer. Attach chamber (SA-1Y) and clamp to the viscometer.
3) Lower the DV-III Ultra into measurement position. Operate the viscometer at 50 or 60 RPM
until the chamber is fully ooded.
4) The viscosity standard uid, together with the spindle, should be immersed in the bath for a
minimum of 1 hour, stirring the uid periodically (operate at 50 or 60 RPM periodically), prior
to taking measurements.
5) After 1 hour, check the temperature of the viscosity standard uid with an accurate
thermometer.
6) If the uid is at test temperature (+/- 0.1°C of the specied temperature, normally 25°C),
measure the viscosity and record the viscometer reading.
Note: The spindle must rotate at least ve (5) times for one minute, whichever
is greater before readings are taken.
7) The viscosity reading should equal the cP value on the viscosity uid standard to within the
combined accuracies of the viscometer and the standard (as discussed in the section entitled,
Interpretation of Calibration Test Results).
Calibration Procedure for Cone/Plate Viscometers:
1) Follow the gap setting procedures in Appendix A for mechanically adjusting the setting of the
cone spindle to the plate.
2) Refer to Appendix A, Table A-1, and determine the correct sample volume required for the
selected spindle.
3) Select a viscosity standard uid that will give viscosity readings between 10% and 100% of full
scale range. Refer to Appendix B for viscosity ranges of cone spindles. Consult with Brookeld
or an authorized dealer to determine which uid is appropriate.
It is best to use a viscosity standard uid that will be close to the maximum viscosity for a given
cone spindle/speed combination.
Example: LVDV-III Viscometer, Cone CP-42, Fluid 10
Having a viscosity of 9.7 cP at 25°C
At 60 RPM, the full scale viscosity range is 10.0 cP. Thus, the Rheometer reading should
be 97% torque and 9.7 cP viscosity ± 0.197 cP (0.1 cP for the rheometer plus 0.97 cP for the
uid). The accuracy is a combination of Rheometer and uid tolerance (refer to Interpretation
of Calibration Test Results).
4) With the viscometer stopped, remove the sample cup and place the viscosity standard uid into
the cup, waiting 10 minutes for temperature equilibrium.